Remote Control
R&S
®
ESRP
671
User Manual 1175.7445.02 ─ 08
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
CALCulate<n>:DELTamarker<m>:X:RELative?
This command queries the x-value of the selected delta marker relative to marker 1 or
to the reference position (for
CALC:DELT:FUNC:FIX:STAT ON
). The command acti-
vates the corresponding delta marker, if necessary.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
CALC:DELT3:X:REL?
Outputs the frequency of delta marker 3 relative to marker 1 or
relative to the reference position.
Usage:
Query only
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
CALCulate<n>:DELTamarker<m>:Y
This command queries the measured value of a delta marker. The corresponding delta
marker is activated, if necessary. The output is always a relative value referred to
marker 1 or to the reference position (reference fixed active).
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
The unit of results depends on the result display and the unit you have selected.
Table 11-19: Base unit
Parameter, measuring function or result display
Output unit
DBM | DBPW | DBUV | DBMV | DBUA
dB (lin/log)
WATT | VOLT | AMPere
dB (lin), % (log)
statistics function (APD or CCDF) on
dimensionless output
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Remote Commands in Receiver Mode