Remote Control
R&S
®
ESRP
647
User Manual 1175.7445.02 ─ 08
Parameters:
<Time>
Range:
10 µs to 100 s
*RST:
1 ms
Example:
SCAN4:TIME 1 ms
Defines a measurement time of 1 ms for the 4th scan range.
[[SENSe:]SWEep:SPACing
<Spacing>
This command selects the frequency step mode.
Note that the command has no effect on the scale and display of the frequency axis.
Parameters:
<Spacing>
LINear
Linear frequency steps with a fix stepsize.
LOGarithmic
Logarithmic frequency steps with the stepsize being a percent-
age of the current frequency.
AUTO
The stepsize is coupled to the resolution bandwidth to get the
best measurement results.
*RST:
LINear
Example:
SWE:SPAC LOG
Selects logarithmic frequency steps.
Manual operation:
See
11.4.6.3
Peak Search
CALCulate<n>:MARKer<m>:PEXCursion
.........................................................................647
CALCulate<n>:PEAKsearch|PSEarch:ADD
...................................................................... 648
CALCulate:PEAKsearch|PSEarch:CLEar[:IMMediate]
........................................................648
CALCulate<n>:PEAKsearch|PSEarch[:IMMediate]
............................................................ 648
CALCulate<n>:PEAKsearch|PSEarch:MARGin
................................................................. 648
CALCulate<n>:PEAKsearch|PSEarch:METHod
................................................................ 649
CALCulate<n>:PEAKsearch|PSEarch:SUBRanges
........................................................... 649
CALCulate<n>:PEAKsearch|PSEarch:SUBRanges:PCOunt
............................................... 650
CALCulate<n>:MARKer<m>:PEXCursion
<Excursion>
This command defines the peak excursion
The peak excursion sets the requirements for a peak to be detected during a peak
search.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Remote Commands in Receiver Mode