Remote Control
R&S
®
ESRP
554
User Manual 1175.7445.02 ─ 08
Table 11-11: Meaning of the bits used in the STATus:QUEStionable:LIMit register
Bit No.
Meaning
0
LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1
LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2
LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3
LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4
LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5
LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6
LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7
LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8 to 14
Unused
15
This bit is always 0.
STATus:QUEStionable:LMARgin Register
This register contains information about the observance of limit margins.
You can read out the register with
STATus:QUEStionable:LMARgin<n>:
or
STATus:QUEStionable:LMARgin<n>[:EVENt]?
Table 11-12: Meaning of the bits used in the STATus:QUEStionable:LMARgin register
Bit No.
Meaning
0
LMARgin 1 FAIL
This bit is set if limit margin 1 is violated.
1
LMARgin 2 FAIL
This bit is set if limit margin 2 is violated.
2
LMARgin 3 FAIL
This bit is set if limit margin 3 is violated.
3
LMARgin 4 FAIL
This bit is set if limit margin 4 is violated.
4
LMARgin 5 FAIL
This bit is set if limit margin 5 is violated.
5
LMARgin 6 FAIL
This bit is set if limit margin 6 is violated.
Remote Control - Basics