System Configuration
R&S
®
ESRP
495
User Manual 1175.7445.02 ─ 08
Hardware Info ← System Info
Opens a dialog box that displays hardware information, e.g. on the frontend and moth-
erboard. Every listed component is described by its serial number, order number,
model information, hardware code, and hardware revision.
Remote command:
VOptions ← System Info
Opens a dialog box that displays a list of hardware and firmware information, including:
Label
Description
Device ID
Unique ID of the device
Instrument firmware
Installed firmware version
BIOS
Installed BIOS version
CPLD
CPLD version
MB-FPGA
Motherboard FPGA version
Data Sheet version
Data sheet version of the basic device
<option>
Installed hardware and firmware options
"Open Source Acknowledgment:
Open"
Displays a PDF file containing information on open source code
used by the R&S
ESRP firmware.
For details on options, refer to
Chapter 3.2.1.10, "Checking the Supplied Options"
Remote command:
System Messages ← System Info
Opens the "System Messages" dialog box that displays the generated system mes-
sages in the order of their occurrence. The most recent messages are placed at the
top of the list. Messages that have occurred since the last display of system messages
menu are marked with an asterisk '*'. The following information is available:
No
device-specific error code
Message
brief description of the message
Component
hardware messages: name of the affected module
software messages: name of the affected software
Date/Time
date and time of the occurrence of the message
If the number of error messages exceeds the capacity of the error buffer, "Message
buffer overflow" is displayed. To delete messages, see
Instrument Setup and Interface Configuration – SETUP Key