Spectrum Measurements
R&S
®
ESRP
329
User Manual 1175.7445.02 ─ 08
Remote command:
Auto Level
Defines the optimal reference level for the current measurement automatically.
The measurement time for automatic leveling can be defined using the
soft-
key.
You can define a threshold that the signal must exceed before the reference level is
adjusted, see
Remote command:
Settings
Opens a submenu to define settings for automatic leveling.
Possible settings are:
●
●
Meas Time Manual ← Settings
Opens an edit dialog box to enter the duration of the level measurement in seconds.
The level measurement is used to determine the optimal reference level automatically
(see the "Auto Level" softkey,
on page 329). The default value is 1 ms.
Remote command:
[SENSe:]ADJust:CONFigure:LEVel:DURation
Meas Time Auto ← Settings
The level measurement is used to determine the optimal reference level automatically
(see the
softkey).
Upper Level Hysteresis ← Settings
Defines an upper threshold the signal must exceed before the reference level is auto-
matically adjusted when the "Auto Level" function is performed.
Remote command:
[SENSe:]ADJust:CONFiguration:HYSTeresis:UPPer
Lower Level Hysteresis ← Settings
Defines a lower threshold the signal must exceed before the reference level is auto-
matically adjusted when the "Auto Level" function is performed.
Remote command:
[SENSe:]ADJust:CONFiguration:HYSTeresis:LOWer
Sweep Type
Opens a submenu to define the sweep type.
Selecting the sweep type is not available for the I/Q analyzer.
Configuration