7.3.9 TRACe Commands
The commands in this section configure and control tracing of the remote interface and
events into a file and enable/disable the display of the SCPI remote trace.
Before you start tracing, configure all settings as desired. Modifying settings while tracing
is active may result in loss of already traced data. Useful exception: Selecting a new target
file while tracing is allowed. For start mode "EXPLicit" a restart of the instrument resets
the settings to the documented default values.
If you want to start tracing already during start of the instrument, configure all settings
(including start mode "AUTO"). Then restart your instrument. Tracing will be started
automatically during the restart, using the already configured settings.
If you use an XML file as trace file, ensure that tracing is stopped properly. If tracing is
aborted instead of stopped, e.g. by shutting down the instrument for stop mode
"EXPLicit", the XML file will be invalid, because some tags are not closed.
When the maximum file size is reached (except for stop mode "BUFFerfull") or if tracing
is started with an already existing trace file, a backup of the trace file is created and the
file itself is reset and overwritten. When the file is full for the second time or tracing is
started the next time, the first backup file is lost because it is overwritten by the next
backup. In order to prevent loss of data, set a sufficient file size, select an appropriate
stop mode and archive/copy completed trace files if you want to keep them.
......................................................................260
.......................................................................261
.....................................................................261
........................................................................262
..........................................................................262
...............................................................262
................................................................263
..........................................................................263
TRACe:REMote:MODE:FILE<inst>:ENABle
<Enable>
Enable or disable tracing of the remote interface to a file for the specified sub-instrument.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<Enable>
ON | OFF
Default value: OFF
Example:
TRACe:REMote:MODE:FILE1:ENABle ON
TRACe:REMote:MODE:FILE2:ENABle OFF
Enable tracing for sub-instrument 1 and disable it for sub-instru-
ment 2.
Firmware/Software:
V2.0.10