![Rohde & Schwarz CMW500 Series Скачать руководство пользователя страница 246](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718246.webp)
Command Reference
R&S
®
CMW500
246
User Manual 1173.9463.02 ─ 02
SYSTem:COMMunicate:NET:GATeway
'<Address>'
Defines the default gateway of the R&S CMW500. This address is valid if DHCP is dis-
abled (
).
Parameters:
'<Address>'
String parameter, gateway consisting of four blocks separated by
dots.
*RST:
a *RST does not affect the LAN service settings
Example:
SYSTem:COMMunicate:NET:GATeway '10.113.0.1'
Select the default gateway.
Firmware/Software:
V1.0.0.4
SYSTem:COMMunicate:NET:DNS:ENABle
<Boolean>
Enables or disables dynamic DNS address configuration.
Parameters:
<Boolean>
-1 | 0
-1:
Dynamic DNS enabled, automatic DNS address setting
0:
Dynamic DNS disabled, manual address setting
*RST:
a *RST does not affect the LAN service settings
Example:
SYSTem:COMMunicate:NET:DNS:ENABle 0
Disable dynamic DNS address configuration, enable manual
address setting.
Firmware/Software:
V1.0.0.4
SYSTem:COMMunicate:NET:DNS
'<Preferred>', '<Alternate>'
Defines the preferred and the alternate DNS addresses of the R&S CMW500. The
addresses are valid if dynamic DNS address configuration is disabled (
).
Parameters:
'<Preferred>'
'<Alternate>'
String parameters, DNS addresses consisting of four blocks sep-
arated by dots.
*RST:
a *RST does not affect the LAN service settings
Example:
SYSTem:COMMunicate:NET:DNS '10.0.2.166',
'10.0.23.159'
Select the DNS addresses.
Firmware/Software:
V1.0.0.4
Instrument-Control Commands