Cable and Antenna Test Mode
R&S
®
Cable Rider ZPH
116
User Manual 1321.0950.02 ─ 03
You can now perform measurements using the temporary cable definition.
See
After you have defined the cable characteristics, you still have to activate the use of the
customized cable.
1. Select the "Cable Config" softkey.
The R&S Cable Rider ZPH opens a submenu to select the cable model.
2. Enable the checkbox to turn on the "User Model" menu item.
The R&S Cable Rider ZPH uses the customized cable model.
7.2.1.4
Working with a DTF List
"DTF Meas Overview", "DTF List" and "DTF Threshold" are only available in the dis-
tance to fault (DTF) measurement.
The DTF list shows the results of the distance to fault measurement in numerical form.
1. Press the "DTF CONFIG" key.
2. Select the "DTF List" softkey.
The R&S Cable Rider ZPH opens a table that lists the peaks it has found during
the measurement.
1 = Shows the peak the results refer to
2 = Shows the distance from the measurement plane to the peak.
3 = Shows the magnitude of the peak.
To limit the information in the list, you can define a new threshold level for the DTF list,
so that only the peaks above a certain level are included in the list.
Configuring Cable and Antenna Tests