Using Power Sensors
R&S
®
AREG100A
71
User Manual 1178.7417.02 ─ 04
"User"
Sets a user defined frequency.
Example:
If you have a frequency converting device between the generator and
the DUT. If the frequency converter doubles the frequency, you can
set twice the frequency in the R&S
AREG100A. The R&S power sen-
sor considers this RF frequency setting.
to the measurement's frequency.
Remote command:
Frequency ← Sensor settings
Defines the frequency value if "Source > User" is used.
Remote command:
Level Offset State,Level Offset ← Sensor settings
Activates and defines a level offset which is considered in the power meassurement
result. The level offset value is always expressed in dB, irrespective of the display of
the measurement result.
This function allows you to consider, for example, an attenuator in the signal path.
Remote command:
:SENSe<ch>[:POWer]:OFFSet:STATe
Filter ← Sensor settings
Selects the way the length of the used filter is defined.
See also
"About the measuring principle, averaging filter, filter length, and achieving
"Auto"
Selects the filter length automatically and adjusts it to the measured
value. The value is indicated with the parameter
.
When high output power is applied, the filter length and therefore the
measurement time can be short.
When low output power is applied, the filter length and therefore the
measurement time is increased which reduces the considered noise
content in your measurement.
"User"
The filter length is defined manually, with the parameter
As the filter length works as a multiplier for the time window, constant
filter length results in a constant measurement time.
Values 1 and 2N are allowed.
NRP Power Viewer