Instrument Functions
R&S
®
NGP800
59
User Manual 5601.5610.02 ─ 04
Trigger in
parameters
Trigger conditions
Description
"Arb Step Group"
Selected channel
function steps to
the next group when the selected logic level
is met.
"Ramp"
Selected channel
function is
enabled when the selected logic level is met.
"Logging"
Selected channel
function is enabled
when the selected logic level is met.
"Statistics"
Selected channel statistics function is
enabled when the selected logic level is met.
"AnalogIn"
Selected channel
is enabled
when the selected logic level is met.
"Inhibit"
Selected channel output is inhibited when
the selected logic level is met.
If the selected channel output is put to inhibit
state, manual or remote operation on
selected channel output is no longer possi-
ble .
To reverse the inhibit state, remove the
source of the trigger signal. You can either
disable the affected DIO interface or remove
the source from the affected DIO interface at
the rear panel.
Active Level
High
Set the logic level of the trigger in signal.
Low
Table 6-2: Trigger out parameters and conditions
Trigger out
parameters
Trigger conditions
Description
Channel
"--", "Ch 1", "Ch 2", "Ch 3", "Ch 4"
Output channel selected to monitor for trig-
ger conditions.
If "--" is selected, no channel is selected for
trigger response.
Mode
Output
Output the selected logic level when the out-
put is turned on at the selected channel.
Fuse
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
See
Chapter 6.4.1, "Overcurrent Protection
Operation mode
●
"CC": Output the selected logic level
when the selected channel operates in
the CC mode. See
●
"CV": Output the selected logic level
when the selected channel operates in
the CV mode. See
Digital Trigger I/O