Preparing for use
R&S
®
ENV216
17
User Manual 5201.6693.14 ─ 05
The device is not equipped with an ON / OFF switch. Instead, the device plug of the
connecting cable serves for disconnection from the AC supply (see
1.
DANGER!
Risk of electric shock. Before you connect the V-network to the power
source, you have to connect the R&S
ENV216 to a protective ground terminal. An
unearthed R&S
ENV216 is live. Touching a live electrical device causes serious
personal injury, or even death.
The protective ground connection must remain until you have disconnected the V-
network from the power supply.
Connect the R&S
ENV216 to the mains power supply.
2. If the phase connection is made correctly, LED L1 (1) must light up.
If it does not:
● For mains plugs with symmetrical grounding connection (e.g. Schuko plugs)
the plug's polarity must be reversed.
● For mains plugs with asymmetrical grounding connection, phase L1 and neutral
N must be swapped by reconnecting.
● For country-specific sockets, see
Chapter A, "Country-specific Connector Mod-
The plug on the mains cable can be adapted using a country-specific adapter.
However, it must be ensured that the adapter has a power rating of at least 16 A.
Assembly must be carried out exclusively by qualified technicians.
Connecting to power