Remote Control
R&S
®
NRT2
133
User Manual 1178.5550.02 ─ 04
11.9
INPut
........................................................................................ 133
.......................................................................................133
INPut<Sensor>:PORT:SOURce[:VALue]
...........................................................................133
INPut<Sensor>:PORT:SOURce:AUTO
............................................................................. 134
INPut<Sensor>:PORT:OFFSet
<offs>
This command allows the transmission loss in a cable connecting the desired mea-
surement point and the sensor to be taken into account. If the measurement position is
set to LOAD, i.e.
INPut<n>:PORT:POSition LOAD
, the cable is assumed to be con-
nected between the sensor and the load, otherwise between the sensor and the
source.
Parameters:
<offs>
Range:
0.0 to 100.0
*RST:
0.0
Default unit: dB
Example:
INP1:PORT:OFFS 1.25
Manual operation:
See
INPut<Sensor>:PORT:POSition
<pos>
This command selects the reference plane. It defines the sensor port to which the
measurement results are to be referred to.
Parameters:
<pos>
SOURce | LOAD
SOURce
Referred to source.
LOAD
Referred to load.
*RST:
SOURce
Example:
INP2:PORT:POS SOUR
Manual operation:
See
INPut<Sensor>:PORT:SOURce[:VALue]
<val>
Defines the forward direction for the given input if the automatic assignment has been
switched off by
INPut{Sensor{1...128}}:PORT:SOURce:AUTO OFF
.
Parameters:
<val>
Range:
1 to 2
*RST:
1
Manual operation:
See
INPut