![Rohde & Schwarz 1424.7805K02 Скачать руководство пользователя страница 331](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791331.webp)
Remote control commands
R&S
®
NRX
331
User Manual 1178.5566.02 ─ 09
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
OFF | ON
*RST:
OFF
13.8.2
Sensor modes
Further information:
●
●
Chapter 13.8.8.1, "NRT mode settings"
CALCulate<Measurement>[:CHANnel<Channel>]:BURSt:TIMing:EXCLude:STARt
CALCulate<Measurement>[:CHANnel<Channel>]:BURSt:TIMing:EXCLude:STOP
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:DCYCle:STATe
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:DCYCle[:VALue]
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:ESAMpling
.................................333
CALCulate<Measurement>[:CHANnel<Channel>]:SAMPling
..............................................333
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:BURSt:DTOLerance
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:APERture[:VALue]
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:SMOothing[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:ESAMpling:AUTO[:STATe]
CALCulate<Measurement>[:CHANnel<Channel>]:BURSt:TIMing:EXCLude:STARt
<value>
Effective for burst average measurements.
Sets the time interval at the beginning of bursts that is excluded from the measure-
ment. Thus, signal overshoots are omitted.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
"Exclude from Start, Exclude from End"
Configuring sensors