Remote Control Commands
R&S
®
NRX
354
User Manual 1178.5566.02 ─ 08
Sensor parameter
Remote control command
Value
Trace capture time
CALCulate<Measurement>:TRACe:X[:SCALe]:LENGth
10.250
ms
Trace number of
points
CALCulate<Measurement>:TRACe:X:POINts
261
Trace realtime
[SENSe<Sensor>:]TRACe:REALtime
OFF
Trigger delay
TRIGger<Measurement>[:CHANnel<Channel>]:
DELay[:VALue]
0
s
Trigger level
TRIGger<Measurement>[:CHANnel<Channel>]:LEVel
-
30
dBm
Trigger source
TRIGger<Measurement>[:CHANnel<Channel>]:
SOURce
INT
Trigger slope
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
POS
Trigger holdoff
TRIGger<Measurement>[:CHANnel<Channel>]:
HOLDoff
19.9
ms
Trigger dropout time
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
0
s
Trigger hysteresis
TRIGger<Measurement>[:CHANnel<Channel>]:
HYSTeresis
3
dB
Table 14-12: Bluetooth DH1
Sensor parameter
Remote control command
Value
Sensor mode
TRACe
Aperture time
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer][:AVG]:APERture[:VALue]
1.25
ms
Duty cycle
CALCulate<Measurement>[:CHANnel<Channel>]:
CORRection:DCYCle[:VALue]
29.28
%
Duty cycle correction
CALCulate<Measurement>[:CHANnel<Channel>]:
CORRection:DCYCle:STATe
ON
Smoothing
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer][:AVG]:SMOothing[:STATe]
OFF
Dropout tolerance
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer]:BURSt:DTOLerance
366
μ
s
Exclude time start
CALCulate<Measurement>:TSLot:TIMing:EXCLude:
STARt
10
μ
s
Exclude time end
CALCulate<Measurement>:TSLot:TIMing:EXCLude:
STOP
15
μ
s
Timegate offset
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:
AVG]:OFFSet[:TIME]
1: 10
μ
s
2: 660
μ
s
Timegate length
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:
AVG]:TIME
1: 341
μ
s
2: 341
μ
s
Timegate fence
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:
AVG][:EXCLude]:MID[:STATe]
OFF
Configuring Sensors
www.allice.de
Allice Messtechnik GmbH