TRIGger<undef>:ALL:DTIMe
<dropout>
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
<dropout>
If the trigger level has been underrun with internal triggering, it is sometimes useful to
wait some time (drop-out time) before retriggering. The measurement of unwanted sig-
nal components can thus be suppressed. See
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<dropout>
Range:
0.0 to 10.0
*RST:
0.0
Default unit: s
Manual operation:
TRIGger<Measurement>[:CHANnel<Channel>]:EXTernal<Port>:IMPedance
<impedance>
Requires a power sensor with a trigger input.
Sets the termination resistance of the external trigger signal that is supplied at the trig-
ger input/output of the power sensor. Choose a setting that fits the impedance of the
trigger source to minimize reflections on the trigger signals.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
<Port>
1 to 2
Power sensor ports; 1 = USB port, 2 = trigger I/O connector
Parameters:
<impedance>
HIGH | LOW
HIGH
~
10
k
Ω
LOW
50
k
Ω
*RST:
HIGH
Manual operation: