Remote Control Commands
R&S
®
NRX
314
User Manual 1178.5566.02 ─ 08
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>:STATistics:PDF:MARKer:Y:POSition
<value>
Positions the y-marker to a measurement value of the PDF statistics measurement.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 10000.0
*RST:
0.0
Default unit: -
Manual operation:
See
14.6.12
NRT
Further information:
●
CALCulate<Measurement>:RELative<DirectionalChannel>:STATe
CALCulate<Measurement>:RELative<DirectionalChannel>:CCDF
......................................315
CALCulate<Measurement>:RELative<DirectionalChannel>:POWer[:MAGNitude]
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:RCOefficient
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:RFRatio
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:RLOSs
........................... 316
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:SWR
..............................317
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio[:MAGNitude]
[SENSe<Sensor>:]FUNCtion:CONCurrent
........................................................................318
[SENSe<Sensor>:]FUNCtion:OFF:ALL<Channel>
.............................................................318
[SENSe<Sensor>:]FUNCtion:OFF[:FUNC]
........................................................................318
[SENSe<Sensor>:]FUNCtion:STATe?
...............................................................................319
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
.......................................................319
[SENSe<Sensor>:]POWer[:POWer]:RANGe:AUTO
........................................................... 319
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit:DETect
............................................. 319
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit:DETect
..................................................319
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit[:STATe]
.............................................320
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit[:STATe]
................................................. 320
[SENSe<Sensor>:]POWer:REFLection:RANGe:LOWer
..................................................... 320
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
..........................................................320
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH