Remote Control Commands
R&S
®
NRX
315
User Manual 1178.5566.02 ─ 08
[SENSe<Sensor>:]POWer:REFLection:RANGe[:UPPer]
.................................................... 320
[SENSe<Sensor>:]POWer[:POWer]:RANGe[:UPPer]
.........................................................320
...................................................................................... 321
CALCulate<Measurement>:RELative<DirectionalChannel>:CCDF
<value>
Sets the reference value for the cumulative distribution function (CCDF).
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
50.0
Default unit: pct
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
on page 56
See
CALCulate<Measurement>:RELative<DirectionalChannel>:POWer[:MAGNitude]
<value>
Sets the reference value for relative ratio measurements.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-120.0 to +150.0
*RST:
+0.0
Default unit: dBm
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
on page 56
See
Measurement Settings and Results