Remote Control Commands
R&S
®
NRX
188
User Manual 1178.5566.02 ─ 08
Parameter list
For continuous average measurements, the following parameters are used.
●
<expected_value>
Optional. Value that is expected for the measurement.
●
<resolution>
Optional. Limit up to which the measurement result should be free of noise.
Corresponds to
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
.
●
<source_list>
Optional. Primary and secondary sensor. The number for each sensor is preceded
by the character
@
. The entire expression is enclosed in parentheses.
Example: (
@
3),(
@
2)
Sensor C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>[:SCALar][:POWer][:AVG]?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
FETCh<Measurement>[:SCALar][:POWer][:AVG]?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
READ<Measurement>[:SCALar][:POWer][:AVG]?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
MEASure<Measurement>[:SCALar][:POWer][:AVG]?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
Measured average power measured by one power sensor.
The
FETCh?
and
READ?
commands send without parameters have a special meaning.
FETCh?
returns the current measured value if it is valid. If a measured value is not yet
available, processing is suspended until a valid result is available.
READ?
starts a measurement and returns a measured value without checking the cur-
rent settings.
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<expected_value_or_source_list>
<numeric> | <expr>
<resolution_or_source_list>
<numeric> | <expr>
<source_list>
<expr>
For the parameter descriptions, see
Usage:
Query only
Making Measurements