Putting into Operation
R&S
®
RT-ZC20B
16
User Manual 1409.8285.02 ─ 04
Tip:
Tap "Save to probe" to store the "Zero adjust" value in the probe box. If
you connect the probe to another channel or to another R&S RTx oscillo-
scope, the value is read out again.
8. If zero adjustment is not possible, turn the coarse adjustment trimmer to bring
the waveform within the range of zero adjustment (see
Risk of instrument damage
When opening the sensor head of the probe, be sure to operate with the
opening lever. If an upper core is forced to open when the sensor head is
locked, the open-close mechanism can be damaged.
3.4
Connecting the Probe to the DUT
1. Check that the system is safe, and that the preparations described in the pre-
ceding sections have been carried out.
2. Pull the sensor slider, so that the clamp opens.
3. Align the sensor so that the current direction indication matches the direction
of current flow through the conductor that you measure. The conductor should
be in the center of the clamp aperture because the measurement may be
affected by the position within the clamp aperture of the conductor being mea-
sured.
4. Press the opening lever until the UNLOCK indication disappears. Make sure
that the sensor head is properly closed.
Connecting the Probe to the DUT