Demagnetizing and Zero Adjustment
Risk of circuit damage
●
Do not demagnetize while the R&S
RT-ZC20 is clamping a conductor to
be measured. Demagnetizing causes current to flow into the conductor,
which may damage parts in the circuit to be measured.
●
Check that the conductor being measured is not clamped when supplying
power to the R&S
RT-ZC20. When power is turned on, a demagnetizing
waveform is initially applied to the output. This is intentional in the design,
and not a fault.
1. With the waveform measuring instrument input at ground, adjust the trace to the
zero position.
2. Connect the R&S
RT-ZC20 current probe and configure the oscilloscope as
.
3. Make sure that there is no conductor in the sensor aperture.