Remote control commands
R&S
®
RTM3000
467
User Manual 1335.9090.02 ─ 09
Parameters:
<Time1>
Center value, maximum value or minimum value depending on
the defined range type.
Range:
20E-9 to 6.87194685440
Increment: Depends on the <Time1> value
*RST:
20E-9
TRIGger:A:WIDTh:DELTa
<Delta>
Defines a variation range around the width value specified using
.
Parameters:
<Delta>
Variation
±Δ
t
Range:
Minimum is 0. Maximum depends on the defined
pulse width (TRIG:A:WIDTH:WITDH).
17.5.5
Video/TV trigger
................................................................................................ 467
TRIGger:A:TV:STANdard
<Standard>
Selects the color television standard.
Parameters:
<Standard>
PAL | NTSC | SECam | PALM | I576 | P720 | P1080 | I1080
PALM = PAL-M
I576 = SDTV 576i (PAL and SECAM)
P720 | P1080 = HDTV 720/1080p (progressive scanning)
I1080 = HDTV 1080i (interlaced scanning)
*RST:
PAL
TRIGger:A:TV:POLarity
<Polarity>
Selects the polarity of the signal. Note that the sync pulse has the opposite polarity.
The edges of the sync pulses are used for triggering,
See also:
Parameters:
<Polarity>
POSitive | NEGative
POSitive
If the video modulation is positive, the sync pulses are negative.
Trigger