Command Reference
R&S
®
ZNA
957
User Manual 1178.6462.02 ─ 12
Setting parameters:
<OldTraceName>
String parameter with old trace name, e.g.
'Trc1'
<NewTraceName>
String parameter with new trace name, e.g.
'S11 Trace'
*RST:
n/a
Example:
See
CONFigure:CHANnel<Ch>:TRACe:REName
Usage:
Setting only
Manual operation:
See
8.3.3
CONTrol Commands
The
Control...
commands allow you to configure the following interfaces:
●
User Port
●
External Handler I/O (Universal Interface, option R&S
ZNBT-Z14)
........................................................................................... 957
CONTrol:AUXiliary:C[:DATA]
<DecValue>
Sets or queries a channel-dependent eight-bit decimal value to control eight independ-
ent output signals at the User Port connector (lines 8, 9, 10, 11 and lines 16, 17, 18,
19). The output signals are 3.3 V TTL signals which can be used to differentiate
between up to 255 independent analyzer states.
CONTrol:AUXiliary:C[:DATA]
itself does not change the analyzer state.
Channel bit definition and activation
The channel bits have the following properties:
●
After a
*RST
of the analyzer all channel bits (including the value for the active,
sweeping channel no. 1) are set to zero; no signal is applied to pins 8 to 11 and 16
to 19 of the User Port connector.
●
The value defined with
CONTrol:AUXiliary:C[:DATA]
is assigned to the
active
).
●
The signals at the User Port connector reflect the channel bits of the
measuring
channel, i.e. the channel for which the analyzer performs a sweep. This channel is
not necessarily identical with the active channel.
●
The signals are switched on as soon as a measurement (sweep) in a channel with
non-zero channel bits is started. They are changed whenever a channel with differ-
ent channel bits becomes the measuring channel.
●
The signals at the User Port connector are maintained after the analyzer enters the
hold state. This happens if all channels use single sweep mode and if all sweep
sequences have been terminated.
●
Pins 16 to 19 may be reserved for monitoring the drive ports 1 to 4 of the analyzer
(
). This leaves up to 16 different monitored channel
states.
SCPI Command Reference