Command Reference
R&S
®
ZNA
921
User Manual 1178.6462.02 ─ 12
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>[:STATe]
<Boolean>
Enables or disables the deembedding function for ground loops. It is allowed to change
the circuit model and its parameters while deembedding is enabled.
Suffix:
<Ch>
.
Channel number.
<group>
Port group (DUT) number.
If multiple port groups are configured (see
ON
, then each port group can
have its own de-/embedding models.
Parameters:
<Boolean>
ON - Deembedding active
OFF - Deembedding inactive
*RST:
OFF
Example:
CALC:TRAN:VNET:GLO:DEEM:TND SL
Select the Shunt L circuit model for deembedding.
CALC:TRAN:VNET:GLO:DEEM:PAR:R SL, 2.2E+3; :
CALC:TRAN:VNET:GLO:DEEM ON
Increase the resistance for the Shunt L circuit model to 2.2 kΩ
and enable deembedding.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C
<CircuitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C?
<CircuitModel>
Specifies the capacitance value C in the different circuit models for ground loop
embedding.
Suffix:
<Ch>
.
Channel number
<group>
Port group (DUT) number.
If multiple port groups are configured (see
ON
, then each port group can
have its own de-/embedding models.
Parameters:
<Capacitance>
Capacitance C for ground loop deembedding.
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F