GUI Reference
R&S
®
ZNA
558
User Manual 1178.6462.02 ─ 12
Access:
> "Ground Loop"
For the "1-Port Data" network type, the required touchstone file can also be selected
from here ("...").
This panel can also be activated by selecting the
to its description for background information, parameters and additional remote com-
mands.
If
are configured and
enabled, the panel displays the available port groups. Select the port group whose
ground loop network shall be configured.
Active
The checkboxes in the "Active" column activate or deactivate the configured ground
loop deembedding/embedding (i.e. adds or removes it to/from the calculation flow)
without changing its parameters.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>[:
STATe]
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>[:STATe]
File Name 1
The ellipsis button in the "File Name 1" column is enabled if a 1-port data
work (
*.s1p
file) is selected.
Note:
The loaded S-parameter trace is stored in the active recall set. Persisted recall
sets contain the full embedding and deembedding data so that they can be transferred
to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:GLOop:DEEMbedding<group>
MMEMory:LOAD:VNETworks<Ch>:GLOop:EMBedding<group>
Offset Embed Softtool