Command Reference
R&S
®
ZNA
898
User Manual 1178.6462.02 ─ 12
CALCulate<Chn>:TRANsform:TIME:STATe
<Boolean>
Determines whether the time domain transformation for trace no.
<Chn>
is enabled.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON - time domain representation active.
OFF - frequency domain representation active.
*RST:
OFF
Example:
*RST; :CALC:TRAN:TIME:STAT?
Reset the instrument, activating a frequency sweep, and query
whether the default trace is displayed in the time domain. The
response is 0.
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:STIMulus
<Type>
Selects the type of stimulus to be simulated in the low pass transformation process.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Type>
IMPulse | STEP
IMPulse - impulse response, in bandpass or lowpass mode.
STEP - step response, only in lowpass mode (a bandpass mode
setting
CALCulate<Chn>:TRANsform:TIME[:TYPE]
BPASs
is automatically changed to lowpass).
*RST:
IMP
Example:
See
CALCulate<Chn>:TRANsform:TIME[:TYPE]
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:STOP
<StopTime>
Defines the stop time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<StopTime>
Stop time of the diagram.
Range:
-99.999999999998 s to +100 s.
Increment: 0.1 ns
*RST:
+4E-009 s
Default unit: s
Example:
See
CALCulate<Chn>:TRANsform:TIME:STARt
SCPI Command Reference