GUI Reference
R&S
®
ZNA
331
User Manual 1178.6462.02 ─ 12
Time domain analysis requires option R&S
ZNA-K2. If this option is not installed, the
"Time Domain" and
Background information
Refer to
Chapter 5.7.2, "Time Domain Analysis"
For a comparison of the different transformation types and windows, and for applica-
tion examples, please also refer to the application note 1EZ44_OE (
).
6.5.5.1
Controls on the Time Domain Tab
The contents of the "Time Domain" tab are also displayed on the "Meas" softtool for
non-frequency converting DUTs.
"Low Pass Settings..." opens the
Time Domain
Selects the time domain representation for the active diagram area. The softkey is
enabled if a linear frequency sweep is active (see
on page 418). The ana-
lyzer automatically quits time domain representation when a different sweep type is
selected.
The time domain results are obtained by transforming the measured frequency sweep
data into the time domain using an appropriate mathematical transformation type and
frequency window ("Impulse Response"). The sweep range and the output power for
the active channel is still displayed below the diagram; the displayed time interval is
shown in a second line:
Traces Softtool