Command Reference
R&S
®
ZNA
892
User Manual 1178.6462.02 ─ 12
CALCulate<Chn>:TRANsform:TIME:STIMulus
..................................................................898
CALCulate<Chn>:TRANsform:TIME:STOP
.......................................................................898
CALCulate<Chn>:TRANsform:TIME:WINDow
...................................................................899
CALCulate<Chn>:TRANsform:TIME:XAXis
.......................................................................899
CALCulate<Chn>:TRANsform:COMPlex
<Result>
Converts S-parameters into converted (matched-circuit) Y-parameters or Z-parameters
and vice versa, assuming that port no. i is terminated with Z
0i
so that the three parame-
ter sets are equivalent and the following formulas apply:
ii
ii
i
ii
S
S
Z
Z
1
1
0
,
,
2
0
0
0
0
j
i
Z
Z
S
Z
Z
Z
j
i
ij
j
i
ij
ii
ii
ii
i
ii
Z
S
S
Z
Y
/
1
1
1
1
0
99
...,
,
1
,
,
,
/
1
2
0
0
0
0
j
i
j
i
Z
Z
Z
S
Z
Z
S
Y
ij
j
i
ij
j
i
ij
ij
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Result>
S | Y | Z
S-parameters, Y-parameters, Z-parameters
Example:
*RST; CALC:PAR:MEAS 'Trc1'", '"Y-S22'
Select the converted admittance Y <-- S22 as measurement
parameter of the default trace.
CALC:TRAN:COMP S
Convert the converted Y-parameter into an S-parameter.
CALCulate<Chn>:TRANsform:IMPedance:RNORmal
<Model>
Selects the theory for the renormalization of port impedances. The selection has an
impact on the conversion formulas for wave quantities and S-parameters.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI Command Reference