Command Reference
R&S
®
ZNA
841
User Manual 1178.6462.02 ─ 12
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE
<BandfilterType>
Selects the bandfilter search mode. In contrast to manual control, bandfilter tracking is
not automatically activated.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
This numeric suffix is ignored and may be set to any value
because the bandfilter search functions always use markers M1
to M4.
Parameters:
<BandfilterType>
BPASs | BSTop | BPRMarker | BSRMarker | BPABsolute |
BSABsolute | NONE
Bandfilter search type:
BPASs
–
Bandpass Search Ref to Max
BSTop
–
Bandstop Search Ref to Max
BPRMarker
–
Bandpass Search Ref to Marker
BSRMarker
–
Bandstop Search Ref to Marker
BPABsolute
–
Bandpass Absolute Level
BSABsolute
–
Bandstop Absolute Level
NONE
–
deactivate bandfilter search, result off
*RST:
NONE
Example:
See
CALCulate<Chn>:MARKer<Mk>:BWIDth
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:CENTer
Sets the center of the sweep range equal to the stimulus value of the marker
<Mk>
on
trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number.
Example:
*RST; :CALC:MARK ON
Create marker 1 in the center of the current sweep range and
assign it to trace no. 1.
CALC:MARK:FUNC:CENT
Leave the sweep range unchanged.
Usage:
Event
Manual operation:
See
"Center = Marker / Start = Marker / Stop = Marker / Span =
SCPI Command Reference