GUI Reference
R&S
®
ZNA
669
User Manual 1178.6462.02 ─ 12
Interpolation is used if the sweep points during the measurement differ from the sweep
points during calibration (state label "Cal int"). Compared to the default linear interpola-
tion, the "Higher Order" (cubic spline) interpolation can slightly slow down the mea-
surement but can provide better results, at least if the measured trace shows no signifi-
cant noise or discontinuities. For the factory system error correction, the analyzer
always uses linear interpolation.
Generally, it is recommended to avoid interpolation whenever possible.
Remote command:
User Interface Tab
Provides general user interface configurations.
Sounds
Switches the sounds generated by the analyzer firmware on or off. Sounds are gener-
ated for certain GUI events, when the analyzer generates a notice, status or warning
message (alarm sounds) or during calibration.
The generation of limit fail beeps for
is not affec-
ted by this setting. If enabled, those beeps are generated even if "Sounds" is
unchecked.
Note:
The R&S
ZNA does not have a built-in audio device and loudspeaker. To hear
these sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
SYSTem:SOUNd:ALARm[:STATe]
SYSTem:SOUNd:STATus[:STATe]
Setup Softtool