GUI Reference
R&S
®
ZNA
549
User Manual 1178.6462.02 ─ 12
Source Coherence On
on the "Source Coherence" softtool tab.
Use Port Match and Calibration
Allows you to select between a raw (unchecked), or an optimized phase control, apply-
ing port match corrections and.calibration (checked, default).
If checked, the phase values are relative to a common zero. Otherwise the phase rela-
tion between the different ports is indetermined.
Remote command:
SOURce<Ch>:CMODe:OPTimized[:STATe]
Defined Phase Coherence Control (table)
The controls in the table allow you to activate and define port specific phase settings.
is unchecked. The settings only take
effect if
Def'd Phase Coherence ← Defined Phase Coherence Control (table)
Defines the individual analyzer sources as coherent or non-coherent.
Remote command:
SOURce<Ch>:CMODe:PORT<Pt>[:STATe]
Phase ← Defined Phase Coherence Control (table)
Phase at the calibratied reference plane of the related port.
Remote command:
SOURce<Ch>:CMODe:PORT<Pt>:PHASe