GUI Reference
R&S
®
ZNA
500
User Manual 1178.6462.02 ─ 12
For channels that are linked to a "Cal Group" (using "Apply" or "Apply to All"), a new
calibration overwrites the cal group data and hence affects all channels that are also
linked to this cal group. An "Overwrite Warning" is displayed in this case. To continue
with the calibration, confirm by using button "Overwrite Current File?" or
Remote command:
MMEMory:STORe:CORRection
MMEMory:LOAD:CORRection
MMEMory:LOAD:CORRection:MERGe
Pool / Delete from Pool
The "Pool" table shows all correction data sets
<CalGroup_name>.cal
in the direc-
tory
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Calibration\Data
.
The name of a pool data set can be modified directly in the corresponding table cell.
"Delete from Pool" deletes a cal group file from the pool.
Note
however, that calibra-
tions being used in any of the opened recall sets cannot be deleted.
Remote command:
Preset User Cal
Selects a cal group from the pool that is activated during a user-defined preset.
Remote command:
Resolve Pool Link / Remove Pool Link
Deletes a link between the selected channel and a "Cal Group" (previously created
using "Apply" or "Apply to All"). With "Resolve Pool Link", the cal group data are still
used as a channel calibration ("Channel Cal") for this channel. With "Remove Pool
Link" the channel calibration is removed.
Remote command:
MMEMory:LOAD:CORRection:RESolve
Ch<n> Calibration Properties/Cal Group <n> Properties
Displays the basic channel settings and the properties of the system error correction
and the power correction for the channel (calibration group) selected in the "Channel
State" ("Pool") table.
Cal Softtool