GUI Reference
R&S
®
ZNA
284
User Manual 1178.6462.02 ─ 12
6.2.11.1
Controls on the Spectrum Tab
The controls on the "Spectrum" tab allow you to select a port and to define how the
spectrum trace is calculated. The sweep parameters and other channel settings are
defined in the
Chapter 6.2.11.2, "Spectrum Setup Dialog"
Spectrum Analysis...
Measurement
Selects the physical VNA port for the spectrum measurement.
Remote command:
CALCulate<Ch>:PARameter:SDEFine 'SA1' | 'SA2' | ...
Trace Mode
Defines how the spectrum trace is smoothed and averaged.
"Clear Write"
Overwrite mode: the trace is overwritten by each sweep.
"AVG"
The pointwise arithmetic mean over several sweeps.
"Min Hold"
The minimum value is determined from several sweeps and dis-
played. The R&S
ZNA saves the sweep result in the trace memory
only if the new value is lower than the previous one.
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
ZNA saves the sweep result in the trace memory
only if the new value is greater than the previous one.
The different trace modes are implemented as a combination of the general purpose
Meas Softtool