Command Reference
R&S
®
ZNA
1357
User Manual 1178.6462.02 ─ 12
<Delay>
Optional entry of delay time or phase for UTHRough standard:
<numeric> - entry of the delay time in ps (for non-dispersive
standards) or of an estimate of the phase at the start frequency
of the sweep in deg (for dispersive standards). See also back-
ground information for
AUTO - the analyzer determines the delay time or phase during
the calibration sweep.
*RST:
AUTO
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE[:DUMMy]
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:METHod
<CalType>
Selects a one-port or two-port calibration type for channel
<Ch>
at ports 1/2.
Tip:
[SENSe<Ch>:]CORRection:COLLect:
to select the calibration type for arbitrary analyzer ports or a multiport
calibration type.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
Parameters:
<CalType>
FRTRans | PFRTrans | FTRans | PFTRans | RTRans |
PRTRans | TOM | TSM | TRM | TRL | TNA | TOSM | ETOM |
ETSM | FOPort1 | FOPort2 | FOPort12 | FOPTport | ROPTport |
REFL1 | REFL2 | REFL12 | TPORt | UOSM
Calibration types, TOM, TRM, TRL, TNA, TOSM, Full One Port,
One Path Two Port, Normalization (REFL1, REFL2 and REFL12
for one-port, FRTRans, FTRans, RTRans, and TPORT for two-
port), TOSM with unknown through.
The numbers in the parameters denote the analyzer ports.
Parameters for two-port calibration types contain no numbers
because the command is only valid for ports 1 and 2.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE[:DUMMy]
[SENSe<Ch>:]CORRection:COLLect:SAVE:DEFault
Generates a set of default system error correction data for the selected ports and cali-
bration type. The default data set corresponds to a test setup which does not introduce
any systematic errors; none of the measurement results acquired previously
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]
) is taken into account.
R&S ZVR/ZVABT Compatible Commands