Command Reference
R&S
®
ZNA
1340
User Manual 1178.6462.02 ─ 12
<PhyPt>
Test port number of the analyzer. This suffix is ignored; the gen-
erator is selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<arg0>
Limit for the generator power.
Range:
Depending on the generator model.
Increment: 0.1-dB
*RST:
The power limit is not changed after a *RST. The
factory setting is 0 dBm.
Default unit: dBm
Example:
See
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:
Manual operation:
See
Generator Ports
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:LLIMit:VALue
<arg0>
Defines a limit for the source power at port no.
<PhyPt>
. The limit must be enabled
explicitly; see example.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the limit applies to all
channels.
<PhyPt>
Test port number of the analyzer.
Parameters:
<arg0>
Limit for the internal source power at port no.
<PhyPt>
.
Range:
-40 dBm to +10 dBm. The exact range depends on
the analyzer model; refer to the data sheet.
Increment: 0.1-dB
*RST:
The power limit is not changed after a *RST. The
factory setting is 0 dBm.
Default unit: dBm
Example:
SOUR:POW2:LLIM:VAL -10; STAT ON
Limit the analyzer source power at port 2 to -10 dBm and enable
the source power limit.
SOUR:POW2:LLIM:DGR ON
Optimize the automatic level control (ALC) for additional connec-
tors (option R&S ZVA-B16).
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02',
'gpib0', '21'
Configure an R&S SME02 generator as external generator no.
1, assigning the name Ext. Gen. 1 and a GPIB address 21.
SOUR:POW3:GEN:LLIM:VAL -10; STAT ON
Limit the power of the external generator port 3 to -10 dBm and
enable the source power limit.
SCPI Command Reference