Command Reference
R&S
®
ZNA
1338
User Manual 1178.6462.02 ─ 12
OPTimized
Use both LOs, if available; see
Chapter 5.7.14, "Internal Second
LO Generator (up to 26.5 GHz)"
DEFault
Use only one LO
*RST:
DEFault
Manual operation:
See
8.3.20.4
Power Config
Access:
[Pwr Bw Avg] > "Power" > "Power Config...".
SOURce<Ch>:POWer<Pt>:ATTenuation:AUTO:VALue?
.................................................. 1338
SOURce<Ch>:POWer<Pt>:ATTenuation:MODE
..............................................................1338
SOURce<Ch>:POWer<Pt>:ATTenuation:POFFset
...........................................................1339
SOURce<Ch>:POWer<Pt>:ATTenuation:AUTO:VALue?
Suffix:
<Ch>
.
<Pt>
Return values:
<automaticSourceAttenuation>
Usage:
Query only
SOURce<Ch>:POWer<Pt>:ATTenuation:MODE
<mode for source attenuation>
Defines how the attenuation factor for the generated wave at test port no.
<PhyPt>
is
set.
Use
SOURce<Ch>:POWer<PhyPt>:ATTenuation
to define a fixed (manual) genera-
tor attenuation factor.
Suffix:
<Ch>
.
Channel number.
<Pt>
Test port number of the analyzer.
Parameters:
<mode for source
attenuation>
AUTO | MANual
AUTO - automatic generator attenuator setting
MANual - manual entry of a fixed attenuation step
LNOise - Low noise mode (for measurements involving external
power meters with a wide bandwidth)
*RST:
AUTO
SCPI Command Reference