Command Reference
R&S
®
ZNA
1334
User Manual 1178.6462.02 ─ 12
Manual operation:
See
[SENSe<Ch>:]FREQuency:MDELay:UTONe
<Upper tone source signal>[, <port
number>]
Selects the source for the upper tone signal that is used for the mixer delay measure-
ment.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Upper tone source
signal>
PORT | GENerator | NONE
Upper tone source:
NONE
– no source selected
PORT
– internal source at port <source_no>
GENerator
– external generator no. <source_no>, as defined in
the System Configuration – External Generators dialog
Range:
1 to port number of the analyzer/number of external
generators
<port number>
Number of the port for the internal source or of the generator (to
be omitted for source NONE)
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
Manual operation:
See
[SENSe<Chn>:]FREQuency:MDELay:MSTate?
Suffix:
<Chn>
.
Usage:
Query only
MMEMory:LOAD:MDAData
<Channel Id>, <Filename or CDELAy>
Loads the known delay values of a calibration mixer, to be used as a reference for a
mixer delay measurement calibration. Mixer delay measurements are controlled using
the
SENSe<Ch>:FREQuency:MDELay...
commands.
Setting parameters:
<Channel Id>
Channel number
<Filename or
CDELAy>
String parameter to specify the name and directory of the loaded
file. The default extension (manual control) for files containing
known delays is
*.csv
, although other extensions are allowed.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
MMEMory:CDIRectory?
.
Example:
See
SCPI Command Reference