Command Reference
R&S
®
ZNA
1335
User Manual 1178.6462.02 ─ 12
Usage:
Setting only
Manual operation:
See
"Constant Cal Dly/Variable Cal Dly"
MMEMory:LOAD:MDCData
<Channel Id>, <Filename>
Loads correction data for a mixer delay measurement calibration from a specified file
and assigns it to a channel with a specified number. Mixer delay measurements are
controlled using the
SENSe<Ch>:FREQuency:MDELay...
commands.
Use
MMEMory:LOAD:MDAData
to load delay values of a calibration mixer stored in a
file.
Setting parameters:
<Channel Id>
Channel number
*RST:
-
<Filename>
String parameter to specify the name and directory of the cali-
bration file to be loaded. The default extension (manual control)
for mixer delay calibration files is *.mcal, although other exten-
sions are allowed. If no path is specified the analyzer searches
the current directory, to be queried with
MMEMory:CDIRectory?
.
Example:
Assume that a mixer delay measurement has been configured
and enabled (see examples for the
SENSe<Ch>:FREQuency:MDELay...
commands), and that a
calibration mixer is connected to the analyzer.
MMEM:LOAD:MDAD 1, 'C:
\Rohde&Schwarz\NWA\Calibration\MixerDelay\MDelay.csv'
Load mixer data for the calibration mixer.
SENS1:FREQ:MDELay:CORR OFF
Disable the current correction (if a correction is on).
SENS1:FREQ:MDEL:ACQ
Take a calibration sweep.
MMEM:STOR:MDCD 1, 'C:
\Rohde&Schwarz\NWA\Calibration\MixerDelay\Mixer1.mcal'
Store the measured calibration data for the calibration mixer to a
file.
For a new mixer delay measurement:
MMEM:LOAD:MDCD 1, 'C:
\Rohde&Schwarz\NWA\Calibration\MixerDelay\Mixer1.mcal'
Load the previously generated mixer delay calibration file and
apply the calibration data to the current mixer delay measure-
ment.
Usage:
Setting only
SCPI Command Reference