Command Reference
R&S
®
ZNA
1264
User Manual 1178.6462.02 ─ 12
<TransCoeff>
Transmission coefficient
Range:
-300 dB to +200 dB
*RST:
n/a
Default unit: dB
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Manual operation:
See
"Frequency / Transm. Coefficients: Insert, Delete, Delete
SOURce:POWer... (Contd.)
The
SOURce:POWer...
commands define the power of the internal signal source.
.............................................................................1264
................................................................................ 1265
.............................................................................. 1265
............................................................................1265
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
...................................... 1266
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:OFFSet
............................................ 1266
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:SLOPe
............................................. 1267
SOURce<Ch>:POWer<PhyPt>:ATTenuation
...................................................................1268
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet
SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet
............................................. 1269
SOURce<Ch>:POWer<Pt>:EATTenuation
...................................................................... 1270
SOURce<Ch>:POWer<Pt>:EATTenuation:MODE
............................................................1270
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:OFFSet
............................................... 1270
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:PERManent[:STATe]
............................. 1271
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:STATe
................................................. 1272
SOURce<Ch>:POWer<PhyPt>:PERManent[:STATe]
....................................................... 1272
SOURce<Ch>:POWer<PhyPt>:REDuce[:STATe]
.............................................................1273
............................................................................1273
........................................................................... 1274
............................................................................ 1274
SOURce:POWer:GENerator:SDELay
<SettlingDelay>
If external generators are used in the measurement, this defines the minimum settling
delay before the external generators are considered to be ready, i.e. the time between
setting the generator to the requested frequency and power and continuing with the
measurement.
This is a global setting.
Parameters:
<SettlingDelay>
Settling delay per sweep point.
Factory default is 10 µs (=
10
-5
s).
Default unit: s
Manual operation:
See
"External Generators > Minimum Settling Delay"
SCPI Command Reference