Parameters:
<Boolean>
Manual operation:
See
SOURce<Ch>:POWer:ALC:PIParameter:CTIMe
<control time>
Defines the (maximum) settling time of the ALC control loop, per sweep point.
Set
SOURce<Ch>:POWer:ALC:PIParameter:ASETtling ON
loop as soon as the convergence criteria are met.
Suffix:
<Ch>
.
Channel number
Parameters:
<control time>
Default unit: s
Manual operation:
See
SOURce<Ch>:POWer<Pt>:ALC:PIParameter:GAIN
<Value>
Defines the proportional gain of the ALC PI controller at source port <Pt> in channel
<Ch>.
This setting only takes effect if
SOURce<Ch>:POWer<Pt>:ALC:PIParameter
to
MANual
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<Value>
Proportional gain factor Kr [dB]
Range:
0 to 100
Increment: 0.001
*RST:
0.4
Manual operation:
See
SOURce<Ch>:POWer<Pt>:ALC:PIParameter:ITIMe
<Alc control loop coefficients
itime>
Defines the integration time of the ALC PI controller at source port <Pt> in channel
<Ch>.
This setting only takes effect if
SOURce<Ch>:POWer<Pt>:ALC:PIParameter
to
MANual
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
SCPI Command Reference