Command Reference
R&S
®
ZNA
1238
User Manual 1178.6462.02 ─ 12
SOURce<Ch>:POWer<Pt>:ALC:COUPle
<Boolean>
Enforces identical ALC settings for the individual ports in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number. This suffix is ignored; the setting applies to all
ports.
Parameters:
<Boolean>
ON (1)
Enforces identical ALC settings
OFF (0)
Enables port-specific ALC settings
Manual operation:
See
SOURce<Ch>:POWer<Pt>:ALC:PIParameter
<Mode>
Enables automatic or manual setting of the ALC PI controller parameters at source port
<Pt> in channel <Ch>.
See
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<Mode>
AUTO | MANual
AUTO
The control loop parameters are calculated by the analyzer firm-
ware.
MANual
The parameters can be set using
*RST:
AUTO
Manual operation:
See
SOURce<Ch>:POWer:ALC:PIParameter:ASETtling
<Boolean>
Enables or disables "Auto Settling" of the ALC control loop.
If enabled, the firmware exits the control loop as soon as the power meets the conver-
gence criteria and doesn't wait until the "Control Time" (
) has elapsed.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference