Command Reference
R&S
®
ZNA
1204
User Manual 1178.6462.02 ─ 12
Parameters:
<Boolean>
ON
–
use independent segment sweep time settings, to be
defined via
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
OFF
–
reset the segment sweep time in all sweep segments to
the segment sweep time for unsegmented sweeps, defined via
. ON will not restore the previous
values.
The parameter is automatically switched to
ON
when a segment
or if the channel settings in a sweep segment
require a sweep time larger than the unsegmented sweep time.
*RST:
OFF
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1.
SEGM:SWE:TIME 0.1
Increase the segment sweep time to 0.1 s.
SEGM:SWE:TIME:CONT OFF
Couple the sweep times in all segments and reset the sweep
time in segment no. 1 to the initial value.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:SUM?
Returns the total duration of the segmented sweep, calculated as the sum of the
sweep times of the individual segments (
[SENSe<Ch>:]SEGMent<Seg>:SWEep:
).
Suffix:
<Ch>
.
Channel number.
<Seg>
Sweep segment number. This suffix is ignored; the command
returns the sum of all segments.
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
Usage:
Query only
Manual operation:
See
8.3.14.24
[SENSe:]SPECtrum...
............................................................................... 1204
[SENSe<Ch>:]SPECtrum:IREJection
............................................................................. 1205
[SENSe<Ch>:]SPECtrum:DETector
<Type>
Defines the detector for the spectrum measurement, i.e. how the measurement result
for a particular sweep point is derived from the related sample values.
SCPI Command Reference