Command Reference
R&S
®
ZNA
1080
User Manual 1178.6462.02 ─ 12
Query parameters:
<OnePort
StandardType>
MOPen | FOPen | MSHort | FSHort | MOSHort | MOSHort1 |
MOSHort2 | MOSHort3 | FOSHort | FOSHort1 | FOSHort2 |
FOSHort3 | MREFlect | FREFlect | MMTCh | FMTCh |
MSMatch | FSMatch
Standard type.
For more information see
<ConnectorType>,
<CalKitName>,
<CalKitLabel>
Together with <StandardType> these parameters fully identify
the related standard (see
.
<PhysPort>
Number of the physical port for which the S-parameter data is
valid. Can be omitted if the data are valid for all ports.
Usage:
Query only
[SENSe:]CORRection:CKIT:<TwoPortStandard>:WLABel:SDATa?
<ConnectorType>, <CalKitName>[, <CalKitLabel>[, <SParameter>,
<PhysPort1>, <PhysPort2>]]
Reads the S-parameter data for a particular
2-port
cal kit standard previously loaded
from Touchstone file using
.
The cal kit is identified by its name and label.
Query parameters:
<TwoPortStandard> MMTHrough | MFTHrough | FFTHrough | MMLine | MMLine1 |
MMLine2 | MMLine3 | MFLine | MFLine1 | MFLine2 | MFLine3 |
FFLine | FFLine1 | FFLine2 | FFLine3 | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork
Standard type.
For more information see
<ConnectorType>,
<CalKitName>,
<CalKitLabel>
Together with <StandardType> these parameters fully identify
the related standard (see
.
<SParameter>
S11 | S12 | S21 | S22
S-parameter of the 2-port standard.
<PhysPort1>,
<PhysPort2>
Numbers of the physical ports for which the S-parameter data is
valid. Can be omitted if the data are valid for all port pairs.
Usage:
Query only
8.3.14.9
[SENSe:]CORRection:METas
Using
[SENSe:]CORRection:METas
commands you can define additional properties
SCPI Command Reference