Command Reference
R&S
®
ZNA
1026
User Manual 1178.6462.02 ─ 12
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt>
<TouchstoneFile>[, <Interchange>]
Loads data from a specified two-port (
*.s2p
) Touchstone file defining a single ended
circuit model for embedding.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:DATA
to load circuit data from the remote client.
The query returns the name of the loaded file.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<TouchstoneFile>
String parameter to specify the name and directory of the loaded
Touchstone file.
If no path is specified the analyzer searches the current direc-
tory, which can be set and queried using
<Interchange>
FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
Example:
CALC:TRAN:VNET:SEND:EMB:TND FIMP
Select the "Serial .s2p" data circuit model.
MMEM:LOAD:VNET:SEND:EMB2 'C:\Users\Public
\Documents\Rohde-Schwarz\ZNA\VNET\Test.s2p'
Load a Touchstone file and assign it to the physical port no. 2.
Manual operation:
See
MMEMory:MDIRectory
<NewDirectory>
Creates a new subdirectory for mass memory storage in an existing directory.
Setting parameters:
<NewDirectory>
String parameter to specify the new directory. Either the full path
or a subdirectory for the current directory (see
SCPI Command Reference