Trying Out the Instrument
R&S
®
FSVA3000/R&S
®
FSV3000
66
Getting Started 1330.8073.02 ─ 07
4. Drag the "Marker Peak List" icon from the evaluation bar to the lower half of
the display to add a new window for the peak list.
5. Close the SmartGrid mode.
6. To obtain a more conclusive peak list that does not contain noise peaks, for
example, define a threshold that is higher than the noise floor:
a) Press the [MKR] key on the front panel.
b) Tap the "Marker Config" softkey in the "Marker" menu.
c) Tap the "Search" tab in the "Marker" dialog box.
d) In the "Threshold" field, enter
-68
dBm
.
e) Tap the "State" box for "Threshold" to activate its use.
Only peaks that are larger than -68
dBm will be included in the peak list.
The marker peak list displays the determined peaks that are above the
defined threshold.
Figure 6-11: Marker Peak List
Displaying a Marker Peak List