Remote Commands
R&S
®
Spectrum Rider FPH
422
User Manual 1321.1011.02 ─ 11
Note: If you use the auto peak detector, the command reads out
positive peak values only.
Example:
TRAC:DATA? TRACE1
Reads out the data for trace 1.
Usage:
Query only
13.6.5
Remote Commands of the Power Meter
The chapter provides information on remote commands that configure and perform
power measurements with the power sensor. These commands are available in power
meter mode only.
Availability of remote commands for Power Sensor measurements
Note that the listed remote commands take effect only if a power sensor is connected.
...........................................................................................422
......................................................................................... 423
Configuring Power Level Readout
........................................................................ 423
........................................................................... 425
................................................................................ 425
.........................................................................................426
............................................................................... 426
Reading Out Measurement Results
......................................................................427
Selecting a Telecommunication Standard
.............................................................427
13.6.5.1
Setting the Frequency
The following chapter describes commands necessary to define frequency settings.
List of commands
.........................................................................................422
CALCulate<1...2>:PMETer:CPOWer:BANDwidth
...............................................................423
[SENSe:]PMETer:FREQuency
<Frequency>
This command sets the frequency of the power sensor.
Parameters:
<Frequency>
Specified in the data sheet.
Default unit: GHz
Example:
PMET:FREQ 500 MHZ
Sets the power sensor's frequency to 500 MHz.
Remote Control - Commands