Remote Commands
R&S
®
Spectrum Rider FPH
311
User Manual 1321.1011.02 ─ 11
13.6.1
Common Commands
The common commands are taken from the IEEE 488.2 (IEC 625-2) standard. A par-
ticular command has the same effect on different devices. The headers of these com-
mands consist of an asterisk "*" followed by three letters. Some of the common com-
mands refer to the
Chapter 13.6.12, "Status Reporting System"
List of Common Commands
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the
EVENt
part of
the
QUEStionable
and the
OPERation
registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE
<Value>
Event status enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value>
Range:
0 to 255
*ESR?
Event status read
Returns the contents of the event status register in decimal form and then sets the reg-
ister to zero.
Return values:
<Contents>
Range:
0 to 255
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH