Remote control basics
R&S
®
NRPxxTWG
95
User Manual 1178.8371.02 ─ 08
USB Resource String
The syntax of the used USB resource string is:
USB::<vendor ID>::<product ID>::<serial number>[::INSTR]
where:
●
<vendor ID> is the vendor ID for Rohde & Schwarz (0x0AAD)
●
<product ID> is the product ID for the Rohde & Schwarz sensor
●
<serial number> is the individual serial number on the rear of the sensor
Table 9-2: R&S
NRPxxTWG USB product IDs
R&S NRPxxTWG power sensor
USB product ID
R&S
NRP75TWG
0x01D1
R&S
NRP90TWG
0x01D2
R&S
NRP110TWG
0x01D3
Example:
USB::0x0AAD::0x01D1::100001
0x0AAD is the vendor ID for Rohde & Schwarz.
0x01D1 is the product ID for the R&S NRP75TWG power sensor.
100001 is the serial number of the particular power sensor.
9.2
Status reporting system
The status reporting system stores all information on the current operating state of the
sensor, and on errors which have occurred. This information is stored in the status reg-
isters and in the error queue. You can query both with the commands of the
STATus
subsystem.
9.2.1
Overview
shows the hierarchical structure of information in the status registers.
Status reporting system