Preparing for use
R&S
®
NRPxxTWG
15
User Manual 1178.8371.02 ─ 08
Setup
1
2
3
4
5
6
Figure 3-1: Setup with an R&S
NRP
‑
ZKU cable
1 = Signal source
2 = R&S NRPxxTWG power sensor
3 = Host interface connector
4 = R&S
NRP
‑
ZKU cable
5 = USB connector
6 = Computer with installed VISA driver or R&S
NRP Toolkit
.
1. Connect the R&S
NRP
‑
ZKU cable to the power sensor. See
the host interface of the power sensor"
2. Connect the R&S
NRP
‑
ZKU cable to the computer.
3.
NOTICE!
Incorrectly connecting or disconnecting the power sensor can damage
the power sensor or lead to erroneous results. Ensure that you connect or discon-
nect the power sensor as described in
Chapter 3.4, "Connecting to a DUT"
on page 12.
Connect the power sensor to the signal source.
4. On the computer, start a software application to view the measurement results.
Chapter 5, "Operating concepts"
3.7.1.2
R&S
NRP
‑
Z5 sensor hub setup
The R&S
NRP
‑
Z5 sensor hub (high-speed USB 2.0) can host up to four R&S
NRPxxTWG power sensors and provides simultaneous external triggering to all con-
nected sensors.
Required equipment
●
1 to 4 R&S NRPxxTWG power sensors
●
1 R&S
NRP
‑
ZK6 cable per sensor
●
R&S
NRP
‑
Z5 sensor hub with external power supply unit and USB cable
●
BNC cables to connect the trigger input and trigger output signals (optional)
Connecting to a controlling host