Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.22
E-4
SENSe:TRACe:POINts[?] 1 to 1024
This command defines the time resolution of the measurement result. Each "point" represents a time
interval whose duration is obtained from the trace length (command
SENSe:TRACe:TIME
) divided by
N-1, where N is the number of "points". The measurement result for a "point" equals the average power
over the associated time interval.
The query yields the number of set "points".
Default setting:
100
SENSe:TRACe:REALtime[?] OFF | ON
SENSe:TRACe:REALtime ON
activates realtime processing for the
Trace
mode so that single events
can also be recorded in this setting. Only one sampling sequence per measurement is recorded. Since
the averaging filter of the sensor is not used, the settings made with
SENSe:TRACe:AVERage:STATe
and
SENSe:TRACe:AVERage:COUNt
are ignored. However, they remain unchanged, and are thus
applied again when realtime processing is deactivated
.
The query yields
•
1
for
OFF,
•
2
for
ON.
Default setting:
OFF
SENSe:TRACe:TIME[?] 0.0001 to 0.3
SENSe:TRACe:TIME
sets the trace length in the
Trace
mode. This time period is divided into a number
of equal intervals, in which the average power is determined. The number of intervals equals the
number of test points, which is set with the command
SENSe:TRACe:POINts.
The query yields the set length in s.
Default setting: