Remote Control - Commands
R&S NRP-Z5x
1137.0463.12
6.6
E-7
SENSe (Sensor Configuration)
The sensor is configured by means of the commands of the groups
SENSe
and
TRIGger
.
Table 6-2
Commands of the
SENSe
system
Command
Parameter
Unit
Remarks
SENSe
:AVERage
:COUNt[?]
1 to 65536
:AUTO[?]
OFF | ON | ONCE
:MTIMe[?]
0.01 to 999.99
s
:NSRatio[?]
0.0 to 1.0
dB
:RESolution[?]
1 to 4
:TYPE[?]
RESolution | NSRatio
:RESet
No query
:STATe[?]
OFF | ON
:TCONtrol[?]
MOVing | REPeat
:CORRection
:DCYCle[?]
0.001 to 99.999
%
:STATe[?]
OFF | ON
:OFFSet[?]
–200.0 to 200.0
dB
:STATe[?]
OFF | ON
:SPDevice:STATe[?]
OFF | ON
:FREQuency[?]
0.0 to 18.0e9 (NRP-Z51) or 33.0e9 (NRP-Z52) or
40.0e9 (NRP-Z55 Mod. 03) or
44.0e9 (NRP-Z55 Mod. 04)
50.0e9 (NRP-Z56) or 67.0e9 (NRP-Z57)
Hz
:FUNCtion[?]
“POWer:AVG“
:POWer
:AVG
:APERture[?]
0.001 to 0.3
s
:BUFFer
:SIZE[?]
1 to 1024
Содержание 1138.0005.03
Страница 20: ......
Страница 30: ...Connecting the sensor R S NRP Z5x 1137 0463 12 1 8 E 7 Fig 1 6 Typical test setup with R S NRP Z5 and PC...
Страница 52: ......
Страница 54: ......
Страница 58: ......